The customer reported a leak, and loose tubing and a broken assembly below the probe wipe of the coulter lh 780 hematology analyzer.The leak was described as bluish reagent, approximately 5 ml in volume and contained within the analyzer.The customer indicated that the leak did not appear to affect sample or reagent integrity, and there was no evidence of cross-contamination.The system was stopped to avoid impact on electrical or optical performance of the instrument.The instrument did not generate an error message.There were no erroneous results either generated or reported.Samples were repeated on a backup analyzer and the results appear to correlate.A beckman coulter field service engineer (fse) was sent to the customer's facility to evaluate the analyzer.The operator was wearing a lab coat and gloves at the time of the incident.There was no report of biohazard exposure to mucous membranes or open wounds.There was no death, injury or affect to user or patient treatment.
|