• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JABIL SINGAPORE KANGAROO EPUMP - NEW; FEEDING PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JABIL SINGAPORE KANGAROO EPUMP - NEW; FEEDING PUMP Back to Search Results
Model Number 382400
Device Problems Thermal Decomposition of Device (1071); Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
It was reported to covidien on (b)(6) 2014 that a customer had an issue with a feeding pump.The customer reported the battery burned and service found melted plastic in the battery bay, external signs of melting.
 
Manufacturer Narrative
Submit date: 10/23/2014.An investigation is currently underway, upon completion, the results will be forwarded.
 
Manufacturer Narrative
The device history record (dhr) was reviewed indicating that product was released accomplishing all quality standard requirements.There were no non-conforming issues reported during the production of this product for a similar condition as reported in this complaint.One kangaroo epump - new was returned to the service department with a specific request for service/repair: the customer states battery burned.Initial inspection of the unit found that there was no battery received with the pump and that there was melted plastics on and around the battery door, identifying a failure with the unit, therefore this report will be considered confirmed.The unit was escalated to a senior technician who found corrosion on the pcba and bent pins on the battery harness that may have contributed to the battery burning.Another potential cause of the battery burning is the battery wires being pinched between the battery door and the back housing.The kangaroo epump was manufactured in 2013.According to the kangaroo epump ifu, this pump meets es60601-1:2005, ul60601-1 standards with respect to electrical shock, fire and mechanical hazards.To assure the safe and reliable operation of this pumping system, do not attempt to operate without an approved battery pack installed in the pump.Use of a battery pack not supplied by covidien can be dangerous and voids all warranties and performance specifications.A review of the service history records indicates there is one other service history recorded for this system.This information will be utilized for trending purposes to determine the need for corrective action.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO EPUMP - NEW
Type of Device
FEEDING PUMP
Manufacturer (Section D)
JABIL SINGAPORE
16 tampines industrial crecsent
singapore
SN 
Manufacturer (Section G)
JABIL SINGAPORE
16 tampines industrial crecsent
singapore
SN  
Manufacturer Contact
elaine bishop
15 hampshire street
mansfield, MA 02048
5084524686
MDR Report Key4197293
MDR Text Key4979915
Report Number3009247326-2014-00025
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number382400
Device Catalogue Number382400
Device Lot NumberC1350030
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received01/19/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-