The device history record (dhr) was reviewed indicating that product was released accomplishing all quality standard requirements.There were no non-conforming issues reported during the production of this product for a similar condition as reported in this complaint.One kangaroo epump - new was returned to the service department with a specific request for service/repair: the customer states battery burned.Initial inspection of the unit found that there was no battery received with the pump and that there was melted plastics on and around the battery door, identifying a failure with the unit, therefore this report will be considered confirmed.The unit was escalated to a senior technician who found corrosion on the pcba and bent pins on the battery harness that may have contributed to the battery burning.Another potential cause of the battery burning is the battery wires being pinched between the battery door and the back housing.The kangaroo epump was manufactured in 2013.According to the kangaroo epump ifu, this pump meets es60601-1:2005, ul60601-1 standards with respect to electrical shock, fire and mechanical hazards.To assure the safe and reliable operation of this pumping system, do not attempt to operate without an approved battery pack installed in the pump.Use of a battery pack not supplied by covidien can be dangerous and voids all warranties and performance specifications.A review of the service history records indicates there is one other service history recorded for this system.This information will be utilized for trending purposes to determine the need for corrective action.
|