• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER COULTER LH 780 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER COULTER LH 780 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 723585
Device Problems Fluid/Blood Leak (1250); Aspiration Issue (2883)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/01/2014
Event Type  malfunction  
Event Description
The customer reported a leak from a coulter lh 780 hematology analyzer.The leak occurred from the bottom of the diluter near the left side.The volume of the leak was five to ten milliliters and was not contained within the instrument.In addition, the instrument generated partial aspiration errors.The instrument operator was wearing gloves and a lab coat at the time of the leak.There were no reports of biohazard exposure to the leak.There were no erroneous results generated and patient treatment was not impacted in connection with this event.
 
Manufacturer Narrative
A beckman coulter (bec) field service engineer (fse) was dispatched to evaluate the instrument.The fse found a hole in tubing at pinch valve vl9a.The fse replaced vl9a and the instrument ran without leaks or error messages.Repairs were verified per established procedures.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 780 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
dung nguyen
250 s. kraemer blvd.
brea, CA 92821
7149614941
MDR Report Key4201009
MDR Text Key4978957
Report Number1061932-2014-02642
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K061616
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/01/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number723585
Other Device ID NumberSW VERSION 1B3
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/01/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-