• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC., CRMD TENDRIL ST; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC., CRMD TENDRIL ST; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1782TC/52
Device Problems Device Sensing Problem (2917); Difficult to Advance (2920)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/12/2011
Event Type  malfunction  
Event Description
It was reported that the stylet could not be inserted into the lead and the lead exhibited sensing anomaly.The lead was not implanted.
 
Manufacturer Narrative
This historical complaint is being filed as part of a retrospective review of complaint files in response to a recent fda inspection.There is no change to the actual performance of the product and this report only represents an enhancement to the reporting criteria going forward.Evaluation description: the reported field event of unable to insert stylet was confirmed in the laboratory.Analysis found that the lead was over torqued which caused the inner insulation to be damaged and a short circuit.The short circuit was the cause of the sensing anomaly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL ST
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view ct.
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view ct.
sylmar CA 91342
Manufacturer Contact
cary lawler
15900 valley view ct.
sylmar, CA 91342
8184932621
MDR Report Key4202147
MDR Text Key4980053
Report Number2017865-2014-04801
Device Sequence Number1
Product Code NVN
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2014
Device Model Number1782TC/52
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer10/22/2011
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date09/12/2011
Device Age5 MO
Event Location Hospital
Date Manufacturer Received09/12/2011
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age80 YR
-
-