Catalog Number B1060-040
Device Problems Deflation Problem (1149); Inflation Problem (1310); Difficult to Remove (1528); Folded (2630)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/02/2014
Event Type  malfunction  
Event Description
It was reported that during a procedure of the left superficial femoral artery (sfa) after device preparation the 6 x 40 mm armada 35 balloon dilatation catheter (bdc) was used for post-dilatation of the implanted unspecified stent, however, the balloon during the inflation attempt at 10 atmosphere (atm) appeared to not fully inflate in the middle portion.The bdc was removed from the anatomy and re-inserted for a second dilatation.The bdc was inflated at 10 atm three times for post-dilatation, but after pulling negative to go to neutral with the indeflator the balloon did not deflate.The indeflator was switched to a 60 cc syringe to facilitate deflation, but the balloon did not deflate in the anatomy.The sheath was pushed forward and the balloon was carefully and completely enclosed in the sheath while inflated without issue.After removal from the anatomy and from the sheath, it was noted that the balloon had deflated flat and was not rewrapped.No additional dilatation was performed to the vessel.There was no reported adverse patient sequela.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: the device was returned for analysis.The reported deflation issue was unable to be confirmed.The flat balloon refold was confirmed.Based on a visual dimensional and functional analysis of the returned device, there is no indication of a product deficiency.A review of the lot history record revealed no non-conformances that would have contributed to the reported event.A query of the electronic complaint handling database revealed no other similar incidents reported from this lot.Based on the reviewed information, no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Type of Device
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key4204207
MDR Text Key4978565
Report Number2024168-2014-07019
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 10/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2017
Device Catalogue NumberB1060-040
Device Lot Number40421G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/02/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
INFLATION: INDEFLATOR
-
-