Model Number 8700-0752-01 |
Device Problems
Break (1069); Positioning Problem (3009)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/06/2014 |
Event Type
malfunction
|
Event Description
|
Complainant alleged that an autopulse® li-ion battery (s/n (b)(4)) would not stay seated in the autopulse® platform.However, when placed into the charger, the battery was able to be seated.It appears that one of the tabs in front of the battery was damaged.There were no issues when another battery was placed into the autopulse® platform.No adverse patient sequelae was reported.No further information was provided.
|
|
Manufacturer Narrative
|
Zoll has not yet received the autopulse battery in complaint for investigation.A supplemental report will be filed if and when the product is returned and investigation has been completed.
|
|
Manufacturer Narrative
|
The autopulse battery (s/n (b)(4)) was returned to the manufacturer and evaluated.The battery would not stay seated in a test autopulse platform.Visual inspection identified that one of the battery's right connector pins was damaged leading to the reported issue.Based on condition of the pin, the damage appears to have occurred during use and handling of the battery in the field.
|
|