Catalog Number 397002-001
Device Problems Improper or Incorrect Procedure or Method (2017); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/01/2014
Event Type  malfunction  
Event Description
The customer reported that the companion 2 driver exhibited a "left pressure incorrect" alarm while supporting a patient.The customer also reported that the companion 2 driver was not connected to hospital wall air at the time of alarm, and that connecting the companion 2 driver to hospital wall air resolved the alarm.There was no patient impact.The patient was subsequently switched to the backup driver.This alleged failure mode poses a low risk to the patient because it did not prevent the driver from performing its life-sustaining functions.The companion 2 driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a supplemental mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
CIRCULATORY ASSIST DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
tucson AZ
Manufacturer Contact
carole marcot, esq. vp
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key4212583
MDR Text Key20272700
Report Number3003761017-2014-00215
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/01/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number397002-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/01/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age29 YR
-
-