• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, CARDIAC RHYTHM MANAGEMENT DIVISI OPTISENSE; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, CARDIAC RHYTHM MANAGEMENT DIVISI OPTISENSE; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1999/46
Device Problems Over-Sensing (1438); High Capture Threshold (3266)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/03/2011
Event Type  malfunction  
Event Description
It was reported that the lead exhibited high pacing threshold and noise.The lead has not implanted.
 
Manufacturer Narrative
All info provided by mfr, no medwatch form was received.This historical complaint is being filed as part of a retrospective review of complaint files in response to a recent fda inspection.There is no change to the actual performance of the product and this report only represents an enhancement to the reporting criteria going forward.The reported field event of high threshold and noise was not confirmed.Analysis was normal.No anomaly was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTISENSE
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, CARDIAC RHYTHM MANAGEMENT DIVISI
15900 valley view ct.
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, CARDIAC RHYTHM MANAGEMENT DIVISI
15900 valley view ct.
sylmar CA 91342
Manufacturer Contact
cary lawler
15900 valley view ct.
sylmar, CA 91342
8184932621
MDR Report Key4212702
MDR Text Key5036311
Report Number2017865-2014-05117
Device Sequence Number1
Product Code DTS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2014
Device Model Number1999/46
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer12/08/2011
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date11/03/2011
Device Age4 MO
Date Manufacturer Received11/03/2011
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age58 YR
-
-