The complaint device was returned to stryker sustainability solutions (sss) for an evaluation.The returned device revealed presence of biological material on the distal tip and minor burrs to the distal tip of the outer shaft.Inspection of the proximal end of the inner part confirmed the reported issue as it was broken.The results of the investigation performed substantiated the reported issue as the inner hub was broken.Since appropriate inspections and testing are in place to ensure that devices are able to connect to the hand piece and function properly prior to leaving stryker sustainability solutions and the package was received open, with evidence of use, potential causes for the broken inner hub were determined to be: during clinical use of the device; the material of the device hub may have been stressed when the device was activated, initiating and propagating the crack which resulted in the returned condition of the device; or an internal om material deformation may have been present in the inner shaft hub which may have contributed to the reported issue.The reported event will continue to be monitored through post-market surveillance.A review of the lot control sheet (lcs) indicated that all required inspections and tests were conducted on all arthroscopic shavers in the lot in question indicating that the instrument was in working condition when released from stryker® sustainability solutions.The instructions for use (ifu) for arthroscopic shavers (as) states the following applicable warning related to this failure mode: "do not allow the arthroscopic shaver to come into contact with staples, clips or any metal object to avoid damage to the blade and possible patient injury." "careful handling of the instrument is necessary to avoid damage or breakage as a result of excessive force." "do not apply excessive pressure or ¿side-load¿ the blade during use.Side-loading does not improve the performance of the instrument, can dull the blade, and/or produce metal particulate.".
|