Model Number 20
Device Problems Inappropriate/Inadequate Shock/Stimulation (1574); Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 03/26/2014
Event Type  malfunction  
Event Description
The customer initially reported that their device had its service indicator illuminated and had logged some event codes.Upon evaluation of the device by physio-control, it was observed that the device was missing the positive portion of the biphasic defibrillation waveform.The defibrillator output energy could be reduced by up to approximately 20% from the selected energy level.There was no report of patient use associated with the reported issue.
 
Manufacturer Narrative
Physio-control evaluated the device and verified the reported failure.Physio replaced the therapy pcb assembly and observed proper device operation through functional and performance testing.The device was then returned to the customer for use.Physio further evaluated the removed therapy pcb assembly and determined that the cause of the reported failure was due to a cracked/open filter, designator fl29.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4215915
MDR Text Key5037986
Report Number3015876-2014-01292
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/06/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2014
Is the Reporter a Health Professional? Yes
Device Age10 YR
Event Location Hospital
Date Manufacturer Received10/06/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/21/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-