• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+; CT STATUS+

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+; CT STATUS+ Back to Search Results
Catalog Number 10379680
Device Problems Break (1069); Nonstandard Device (1420); Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/11/2014
Event Type  malfunction  
Event Description
Customer reported that they received broken power supply.Customer reported that instrument lost power during medical examination.Customer also reported that power adaptor was opened and the electrical parts were in contact.There was no report of injury due to this event.
 
Manufacturer Narrative
Customer indicated that they replaced power supply and since then they have had no issue with power supply.The original equipment manufacturer (oem) of these power supplies has identified two root causes for the damaged power supply adaptors.1.The incorrect torque setting was selected by the production operator for the screws used to secure the power supply adaptor housings together.2.Mineral oil was found on the screws used to secure the housings together.The oil is used during the machining process during the production of the screws and was not removed by the screw supplier.These two root causes resulted in a combination of mechanical and chemical stress which led to the cracking of the screw pillars inside the housing top of the power supply adaptor.Siemens issued an urgent field safety notice (31983 rev.A) to notify all affected siemens customers.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+
Type of Device
CT STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key4223434
MDR Text Key5076041
Report Number1217157-2014-00168
Device Sequence Number1
Product Code JIL
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date 10/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number10379680
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/14/2014
Is the Device Single Use? No
Type of Device Usage Unknown
Removal/Correction Number1217157-09-22-2014-001-R
Patient Sequence Number1
-
-