IRIS INTERNATIONAL ICHEM VELOCITY; AUTOMATED URINE CHEMISTRY ANALYZER Back to Search Results
Catalog Number 700-7177-001
Device Problems Charred (1086); Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/06/2014
Event Type  Injury  
Event Description
Customer reported burning smell from the unit.
 
Manufacturer Narrative
A large noise and then a burning smell from the velocity was reported.An iris field service engineer (fse) observed the burnt smell from inside the analyzer caused by the burnt components from the power supply.Fse replaced the power supply and the issue was resolved.There were no reports of injuries, visible flames, smoke emission, and the fire department was not called.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICHEM VELOCITY
Type of Device
AUTOMATED URINE CHEMISTRY ANALYZER
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave.
chatsworth CA 91311
Manufacturer Contact
sudha gupta
9172 eton avenue
chatsworth, CA 91311
8185277272
MDR Report Key4223499
MDR Text Key5063934
Report Number2023446-2014-00173
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/06/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number700-7177-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/24/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-