Model Number 5816
Device Problems Difficult to Interrogate (1331); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/09/2014
Event Type  malfunction  
Event Description
It was reported that pulse generator interrogation yielded an error message.A second attempt yielded no data.A firmware download restored the device to normal functionality, allowing for successful interrogation.No patient symptoms were reported.
 
Manufacturer Narrative
All information provided by manufacturer, no medwatch form was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VICTORY XL DR
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jennie hamel
15900 valley view court
sylmar, CA 91342
8184933462
MDR Report Key4231127
MDR Text Key5001751
Report Number2017865-2014-17999
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2008
Device Model Number5816
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/09/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/18/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
-
-