It was reported that the recently implanted vns patient¿s device was tested during an office visit on (b)(6) 2014 and system diagnostic results revealed high impedance (impedance value = 10,000 ohms).System diagnostic showed lead impedance within normal limits during implant surgery on (b)(6) 2014 (impedance values ¿ 1653 ohms and 1585 ohms).The patient was admitted to the hospital on (b)(6) 2014 and the patient¿s device showed lead impedance within normal limits (impedance values ¿ 4632 ohms, 4306 ohms, and 4197 ohms).On (b)(6) 2014, the patient underwent surgery in which incomplete lead pin insertion was confirmed.After the pin was reinserted, system and generator diagnostic results showed normal device function.
|