It was reported that a patient underwent an atrial flutter (afl) procedure with a c3 interface cable ¿ therapeutic, and a bad no ecg all channels ¿ bs and ic issue occurred.It was reported that the stockert 70 system was clicking and the temperature was going up and down.All ecg¿s were noisy.The noise started when the redel cable was plugged into the patient interface unit (piu).The redel cable and the ablation adapter were replaced without resolution.There was no catheter plugged into the ablation cable that is plugged into the piu.When the c3 interface cable-therapeutic was removed, the noise went away and there was no longer any temperature on the stockert, as it should be.The c3 interface cable-therapeutic was reprocessed many times.A new cable was plugged in and the issue was resolved.The procedure was continued without patient consequence.Upon request, additional information was provided on the event.The noise was observed on all the body surface (bs) and intracardiac (ic) channels on both the carto system and the pruka recording system and there was no signal available to monitor the patient¿s heart rhythm.The issue was a bent pin in the ablation socket.It was in contact with the pin next to it.Once it was replaced, the issue was resolved.The temperature issue was found before use.No energy could be delivered since there was no ablation catheter connected to the stockert generator.Values were not checked at the time of the issue as there was no ablation catheter.The stockert generator was programmed with the default values for the 8mm catheter.Since the noise affected all ic and bs signals on both the carto system and the pruka recording system and there was no signal available to monitor the patient¿s heart rhythm, this event is indicative of a reportable event.
|