• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS BATT F/BPL 14.4VDC; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS BATT F/BPL 14.4VDC; MOTOR, DRILL, PNEUMATIC Back to Search Results
Catalog Number 530.620
Device Problems Melted (1385); Overheating of Device (1437); Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/22/2014
Event Type  malfunction  
Event Description
This is report 2 of 2 for the same event.It was reported from (b)(6) that during an unspecified surgical procedure, it was discovered that the battery reamer drill device overheated and stopped functioning.According to the reporter, the device was then switched off and on again.However, the device was still not functioning.The reporter indicated that the battery inside the casing was checked and it was found that part of the battery housing was melted.There were no delays to the planned surgical procedure as a spare device was available for use.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the control unit was not functioning.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available; a supplemental medwatch report will be sent accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATT F/BPL 14.4VDC
Type of Device
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key4240642
MDR Text Key17696933
Report Number3009450871-2014-10595
Device Sequence Number1
Product Code HBB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK971544
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,company representati
Reporter Occupation Other
Type of Report Initial
Report Date 10/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.620
Device Lot NumberD110325
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received09/30/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/10/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-