Customer reported aspiration errors were obtained after performing a calibration when using the coulter lh 750 hematology analyzer.The customer identified the tubing was broken off the needle assembly which resulted in a leak of about 10 ml of green colored fluid onto the counter.A beckman coulter field service engineer (fse) was sent to the customer's facility to evaluate the analyzer.While correcting the issue associated with the leak, the fse identified the backwash function of the needle assembly was sluggish, and identified a clog in the solenoid 61, which was unrelated to the reported leak.The operator was wearing personal protective equipment of lab coat and gloves when the leak was identified.There was no report of exposure to mucous membranes or cuts.There was no impact to patient results, as patient samples were not tested on the analyzer.There was no death, injury or affect to user or patient treatment.
|
On (b)(4) 2014, a beckman coulter field service engineer (fse) evaluated the analyzer and removed solenoid 61, and found obstruction most consistent with diluent salts in solenoid 61.Cleared obstruction from solenoid 61, and reinstalled the solenoid.The fse retested solenoid 6, and it opened and closed as expected.Instrument performance was verified.(b)(4).
|