• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 25-40-45
Device Problem Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problem No Patient involvement (2648)
Event Date 08/28/2014
Event Type  Other  
Event Description
Sorin group (b)(4) received a report that the oxygen value of the s5 gas blender was out of specification during a preventive maintenance.There was no patient involvement.
 
Manufacturer Narrative
Sorin group (b)(4) manufacturers the s5 gas blender system.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Manufacturer Narrative
(b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the oxygen value of the s5 gas blender was out of specification.During a preventive maintenance.The gas blender was sent to sorin group (b)(4) for evaluation.The reported issue could not be reproduced.The unit was calibrated and tested.A test run for 24 hours with different interventions was performed.During this time no deviations or errors were found.No nonconformities were noted during manufacturing record review.The issue will be monitored for trends and if identified, corrections will be recommended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 GAS BLENDER SYSTEM
Type of Device
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich D 809 39
GM  D 80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich D 809 39
GM   D 80939
Manufacturer Contact
cheri voorhees, mgr.
14401 west 65th way
arvada, CO 80004
3034676527
MDR Report Key4254823
MDR Text Key5044469
Report Number1718850-2014-00420
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K101046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Unknown
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/07/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25-40-45
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/06/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-