MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 97714
Device Problem Device Difficult to Program or Calibrate (1496)
Patient Problems Therapeutic Effects, Unexpected (2099); Complaint, Ill-Defined (2331); Electric Shock (2554)
Event Date 10/31/2014
Event Type  malfunction  
Event Description
The patient noticed approximately two months that the stimulation was not reaching her right foot like it should.Some reprogramming was performed on (b)(6) 2014, and since then the patient has been experiencing ¿zinging¿ shocking ¿out of the blue¿ where her legs go out from under her.The ¿zinging¿ occurred since reprogramming so the next day the patient turned the implantable neurostimulator (ins) off; she does not get shocked with it off.The patient also reported being ¿zapped¿ out of the blue several times while sleeping and standing.The patient was told to keep it on program a only and she did what she was told with the settings, but she was still getting shocked.An information request was made regarding the patient programmer (pp).On (b)(6) 2014, the patient needed to be reprogrammed again because the adaptivestim was not programmed correctly.A week later the patient went to the office and the physician turned the adaptivestim off.Additional information has been requested to find out if any intervention or troubleshooting was required and to obtain the outcome of this event.If additional information is received, a follow up will be sent.
 
Manufacturer Narrative
Concomitant products: product id: 97754, serial# (b)(4), product type: recharger.Product id: 97740, serial# (b)(4), product type: programmer, patient.Product id: 977a260, serial# (b)(4), implanted: (b)(6) 2014, product type: lead.Product id: 3550-29, lot# n457659, implanted: (b)(6) 2014, product type: accessory.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key4255626
MDR Text Key16781096
Report Number3004209178-2014-21688
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/31/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2015
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/31/2014
Date Device Manufactured03/05/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age00054 YR
-
-