Catalog Number 26-1221
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/04/2014
Event Type  malfunction  
Event Description
The product could not cut well during surgery.Another product was used to complete the case.There were no adverse consequences to the patient.
 
Manufacturer Narrative
(b)(4).Upon completion of the investigation, a follow up report will be filed.
 
Manufacturer Narrative
Upon completion of the investigation it was noted that the perforator was evaluated by the supplier and it was found to be within specifications.This perforator met all visual and functional testing requirements.The root cause of the reported problem the product could not cut well during surgery" was not determined.All evaluation tests and inspections had acceptable results.The perforator met functional test method acceptance requirements; proper engagement was achieved with every drilled hole, and there was no premature disengagement.A review of the device history records did not reveal any anomalies.We will continue to monitor for this or similar complaints for this product code.At the present time this complaint is closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOSABLE PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACC.
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key4255913
MDR Text Key5047380
Report Number1226348-2014-12101
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number26-1221
Device Lot NumberGG010S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received02/04/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-