The user facility reported that there was a rough ridge noted approximately 3.5 inches from the distal end, and during a recent bronchoscopy the user experienced difficulty removing the bronchoscope from the tracheostomy tube.Olympus was further informed that after the bronchoscopy, a black rubber piece from the distal end of the bronchoscope was noted in the collected bronchoalveolar lavage (bal) sample.The user facility reported that no device fragment was found inside the patient.There was no patient injury reported.
|
The bronchoscope reference in this report was returned to olympus for evaluation.The evaluation confirmed that the distal end had a rough edge and discoloration on both side of the bending section cover.A portion of the bending section cover was missing, and there were cracks noted.There were excessive dents and scratches noted on the distal end of the bronchoscope.The bronchoscope failed insulation testing due to the damaged distal end cover.In addition, there were scrape marks noted inside the instrument channel 3.5 mm from the distal end.The device was refurbished.The damage found is consistent with physical damage, user techniques, and inaccurate sizing of the endotracheal tube used with the bronchoscope.See scanned page.Also, an olympus endoscopy support specialist (ess) was dispatched to provide the user facility additional training as needed.
|