STRYKER INSTRUMENTS-KALAMAZOO SYSTEM 5 DUAL TRIGGER ROTARY HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
Catalog Number 4205000000 |
Device Problem
Unintended Power Up (1162)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/23/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that during testing conducted at the manufacturer facility, the device ran without user activation after the trigger was released.No medical intervention and no adverse consequences were reported with this event.As this event occurred during testing at the manufacturer facility, there was no patient involvement and no delay to a surgical procedure.
|
|
Manufacturer Narrative
|
Upon disassembly, a displaced straight pin was found, which can lead to the reported event.
|
|