• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM DRIVER Back to Search Results
Catalog Number 595000-001
Device Problems False Alarm (1013); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/26/2014
Event Type  malfunction  
Event Description
The customer reported that the freedom driver exhibited an irreversible fault alarm while supporting a patient.The küster also reported that the patient was rotating the onboard batteries in the freedom driver before going to bed.She inserted the onboard battery backwards, realized her mistake, and then turned the battery around, but was unsuccessful at resolving the fault alarm.The patient was plugged into ac power at the time of alarm.The patient was subsequently switched to the backup freedom driver without adverse impact.Although the freedom driver exhibited an irreversible fault alarm, the driver continued to function as intended.This alleged failure mode poses a low risk to the patient because it does not prevent the freedom driver from performing its life-sustaining functions.The freedom driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a supplemental mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
tucson AZ
Manufacturer Contact
carole marcot, esq., vp
1992 e. silverlake rd.
tucson, AZ 85713
5205451234
MDR Report Key4265035
MDR Text Key5007405
Report Number3003761017-2014-00210
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/26/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/22/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number595000-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/26/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age53 YR
-
-