Shortly after arriving on shift i noted a wet area on the floor beneath 2 medfusion 3500 iv syringe pumps.There were two pumps infusing dextrose 25% into a double lumen uvc at 9ml/hr or 4.5 ml/hr per each pump.I discovered a leak in the iv tubing on one of the pumps just above the in-line filter.I stopped the pump immediately.Increased the rate in the other pump to 9ml/hr.I notified the charge nurse and the nurse practitioner.We changed out the tubing and checked a blood glucose level which was within a normal range (66).The tubing with the leak will be turned in to the nurse manager.Im not sure how it could be prevented unless the leak is noted while priming the tubing.The infusion iv tubing was changed.No further leaks noted after replacement of tubing and the nurse practitioner has been advised.
|