• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROSURGICAL TECHNOLOGY, INC. PACKER/CHANG IOL CUTTER; OPHTHALMIC SCISSORE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROSURGICAL TECHNOLOGY, INC. PACKER/CHANG IOL CUTTER; OPHTHALMIC SCISSORE Back to Search Results
Model Number DFH-0012
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
The surgeon reported that while using a packer/chang iol scissor to cut an iol the scissor blade broke.The broken piece was removed from the eye and there was no impact to the patient.
 
Manufacturer Narrative
The device was returned with corrosion and pitting at the site of the fracture which indicates the device was not cleaned and maintained adequately.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKER/CHANG IOL CUTTER
Type of Device
OPHTHALMIC SCISSORE
Manufacturer (Section D)
MICROSURGICAL TECHNOLOGY, INC.
redmond WA
Manufacturer Contact
8415 154th ave se
redmond, WA 98052
4255560544
MDR Report Key4268537
MDR Text Key5297294
Report Number3019924-2014-00049
Device Sequence Number1
Product Code HNF
Combination Product (y/n)N
Reporter Country CodeBR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial
Report Date 10/30/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDFH-0012
Device Catalogue NumberDFH-0012
Device Lot Number043408
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer10/23/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/20/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-