KARL STORZ GMBH & CO. KG BIPOLAR ELECTRODES; BIPOLAR CUTTING LOOP Back to Search Results
Model Number 27040GP1
Device Problems Thermal Decomposition of Device (1071); Charred (1086); Arcing of Electrodes (2289)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/15/2014
Event Type  malfunction  
Event Description
Allegedly, the doctor was performing a turp when he activated electrodes, 7 arced and broke inside of pt; no material fell into pt.He replaced the electrodes and went on to complete procedure.There was no injury to pt; pt condition post-op was good.
 
Manufacturer Narrative
The esu (autocon) that provides power to these electrodes was over activated.The duty cycle of 25% (10 seconds on/30 seconds off) was not adhered to which caused damage to the generator resulting in a surge which damaged the electrodes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOLAR ELECTRODES
Type of Device
BIPOLAR CUTTING LOOP
Manufacturer (Section D)
KARL STORZ GMBH & CO. KG
tuttlingen
GM 
Manufacturer (Section G)
KARL STORZ GMBH & CO. KG
mittelstrasse 8
tuttlingen 7850 3
GM   78503
Manufacturer Contact
susie chen
2151 e. grand ave.
el segundo, CA 90245-5017
4242188201
MDR Report Key4268639
MDR Text Key5298840
Report Number9610617-2014-00056
Device Sequence Number1
Product Code FAS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061541
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number27040GP1
Device Catalogue NumberSAME
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer10/02/2014
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
-
-