• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, CARDIAC RHYTHM MANAGEMENT DIVISI ZEPHYR XL DR; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, CARDIAC RHYTHM MANAGEMENT DIVISI ZEPHYR XL DR; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number 5826
Device Problems No Device Output (1435); Pacemaker Found in Back-Up Mode (1440)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/13/2012
Event Type  malfunction  
Event Description
It was reported that during an upgrade the device was exposed to cautery and reverted to backup mode, causing loss of output.A device download resolved the issue and the device remained implanted, programmed to sense only.
 
Manufacturer Narrative
All info provided by mfr, no medwatch form was received.This historical complaint is being filed as part of a retrospective review of complaint files in response to a recent fda inspection.There is no change to the actual performance of the product and this report only represents an enhancement to the reporting criteria going forward.Analysis was normal.No anomaly was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEPHYR XL DR
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, CARDIAC RHYTHM MANAGEMENT DIVISI
15900 valley view ct.
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, CARDIAC RHYTHM MANAGEMENT DIVISI
15900 valley view ct.
sylmar CA 91342
Manufacturer Contact
cary lawler
15900 valley view ct.
sylmar, CA 91342
8184932621
MDR Report Key4269707
MDR Text Key5296317
Report Number2017865-2014-03737
Device Sequence Number1
Product Code DXY
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2009
Device Model Number5826
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date06/13/2012
Device Age55 MO
Event Location Hospital
Date Manufacturer Received06/13/2012
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age82 YR
-
-