Catalog Number 700-3320
Device Problems Charred (1086); Device Emits Odor (1425); Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/16/2012
Event Type  malfunction  
Event Description
Customer reported a burning smell from the iq200 unit.
 
Manufacturer Narrative
Customer reported burning smell coming from the iq200.Fse determined that the vertical motor on the spa had burned out.Fse replaced and calibrated spa and completed iq performance verification checklist.Customer ran specimens without any issues.System was operational.There were no reports of injuries, smoke emission, visible flames, and fire department was not called.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IQ200
Type of Device
AUTOMATED URINE MICROSCOPY ANALYZER
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave.
chatsworth CA 91311
Manufacturer Contact
sudha gupta
9172 eton ave.
chatsworth, CA 91311
8185277272
MDR Report Key4272811
MDR Text Key5031831
Report Number2023446-2014-00161
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022774
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/16/2012
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/08/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number700-3320
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/16/2012
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-