• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; CARDIAC OCCLUSION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; CARDIAC OCCLUSION DEVICE Back to Search Results
Model Number 9-ASD-022
Device Problems Leak/Splash (1354); Migration or Expulsion of Device (1395)
Patient Problem No Information (3190)
Event Date 10/30/2014
Event Type  Injury  
Event Description
The atrial septal defect (asd) measured 15 x 19mm on echo and balloon-sized to 22.3mm with good rims.A 19mm amplatzer septal occluder (aso) device was selected but pulled through to the right atrium.A 22mm aso was then selected, position was confirmed on echo and the device was released.Post-procedure day one, a tee revealed a small, residual leak.The physician suspected the aso had migrated and the patient was to undergo surgery to explant the aso and surgically repair the asd.
 
Manufacturer Narrative
(b)(4).Sjm could not evaluate the product involved in this event since it was not returned to us.The device history record for this product was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with sjm specifications and procedures.The results of this investigation are inconclusive because the aso was not returned for evaluation.There was no evidence to suggest there was an intrinsic defect in the occluder, and the cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key4274185
MDR Text Key16558730
Report Number2135147-2014-00107
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
PP000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/04/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/24/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date08/01/2016
Device Model Number9-ASD-022
Device Catalogue Number9-ASD-022
Device Lot Number1108030067
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/04/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/03/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-