• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, DUAL 4 CHANNEL 60CM; SCS EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, DUAL 4 CHANNEL 60CM; SCS EXTENSION Back to Search Results
Model Number 3346
Device Problem Corroded (1131)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/03/2014
Event Type  Injury  
Event Description
Device 2 of 2.Reference mfr.Report# 1627487-2014-20449.
 
Manufacturer Narrative
Udi (di): (b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Device 2 of 2.Reference mfr.Report# 1627487-2014-20449.
 
Manufacturer Narrative
Result: the complaint of ¿corrosion¿ was not confirmed.As received, microscopic inspection revealed broken wires in lead extension ¿a¿ 43 cm away from the stimulation end.However, the outer tubing was not breached at the fracture site.The fracture was consistent with an overstress condition the lead was subjected to while it was inside the patient.As received, lead extension ¿b¿ was clear; both the header and terminal ends were in good condition, and no broken wires or kinks in the lead body were observed.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, DUAL 4 CHANNEL 60CM
Type of Device
SCS EXTENSION
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano, tx
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4274637
MDR Text Key5228464
Report Number1627487-2014-20450
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PP010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/24/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2015
Device Model Number3346
Device Lot Number4194321
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/12/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received01/07/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/08/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 3788, SCS IPG; MODEL 3169(2), SCS LEAD; MODEL 3166(2), SCS LEAD
Patient Outcome(s) Other;
Patient Age58 YR
-
-