Model Number 3700
Device Problem Device Slipped (1584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/24/2014
Event Type  malfunction  
Event Description
Hill-rom received a report from the account stating the brakes were not holding.The bed was located in labor and delivery at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Manufacturer Narrative
Hill-rom technical support suggested to replace the detent.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2011.It is unknown if the facility performed any other preventative maintenance on this bed.The account replaced the detent to resolve the issue.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY 4 BED
Type of Device
BIRTHING BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key4287214
MDR Text Key5296123
Report Number1824206-2014-02679
Device Sequence Number1
Product Code HDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K915779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3700
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/24/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-