Method: the complaint mr290v vented autofeed humidification chamber was returned to fisher & paykel healthcare in (b)(4) and was visually inspected.Results: visual inspection revealed that one of the chamber ports was deformed.The mr290v chamber was set up with an mr850 humidifier and humming v ventilator.The customer has confirmed that the power to the flow source was turned off whilst the humidifier was kept on.The setup also included an impedance valve on one of the chamber ports.A lot check revealed no other similar complaints for lot number 140305.Conclusion: the deformation of the port most likely occurred due to the flow source being turned off whilst the humidifier was kept on.The user instructions were not followed as the humidifier should not be turned on when there is no gas flow.No air flow through the chamber results in excessive heat inside the chamber which may cause the ports to deform.The weight of the impedance valve will also have contributed to the deformation of the port.As part of our manufacturing process, every mr290 chamber undergoes pressure testing for potential leaks and those that fail are rejected.In addition, a visual inspection is performed after the port caps are assembled on to the chamber on the production line.Any product which fails this visual inspection is rejected.The user instructions that accompany the mr850 humidifier state: - if gas flow is interrupted turn the humidifier off.
|