It was reported that the customer smelled something burning from the instrument; however, the exterior of the instrument was not hot to the touch.The instrument's drawer was open.The customer tried to close the drawer by going thru the software as if running a panel and 5 to 10 minutes later observed possibly white smoke emitting from the instrument through the drawer opening.The instrument was not powered but the power cord was disconnected from the electrical source.Although the smoking issue did not set off any fire alarms, a code red was issued to indicate an emergency.It was also reported that the smoke did not fill the room but it aggravated the user's existing cough.Medical intervention was not required.The hospital's biomed personnel removed the instrument from the lab.
|