• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 629029
Device Problems Thermal Decomposition of Device (1071); Device Emits Odor (1425); Failure to Power Up (1476); Smoking (1585)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/07/2014
Event Type  malfunction  
Event Description
While servicing a unicel dxh 800 coulter cellular analysis system in a customer's facility on (b)(6) 2014, a beckman coulter field service engineer (fse) replaced the power supply to address a noisy fan, initially reported by the customer on (b)(6) 2014.After replacing the power supply, the analyzer would not function, i.E.Would not power on.The fse observed smoke from the top of the analyzer, and a burnt odor.The fse identified a burned out chip on the card cage as the cause of the smoke and burnt odor.The fse acquired another power supply and repaired the analyzer on (b)(6) 2014.There were no arcs, sparks of flames observed for this event and the fire department was not called.There was no death, injury, or change to patient treatment or user safety attributed to or connected with this event.
 
Manufacturer Narrative
On (b)(4) 2014, a beckman coulter field service engineer (fse) evaluated the analyzer and replaced the electrical power supply and the card cage resolving the reported issue.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
dung nguyen
250 s kraemer blvd
brea, CA 92821
7149614941
MDR Report Key4301955
MDR Text Key5268817
Report Number1061932-2014-02948
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081930
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/07/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/05/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Service Personnel
Device Catalogue Number629029
Other Device ID NumberSOFTWARE VERSION: 2.0.2.1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/07/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-