It was reported by the sales rep the fms vue was being used in a rotator cuff procedure with a patient that had heavy bleeding and reported there were inflow and outflow issues, where air bubbles were getting into the cavity during the procedure, as well as cannula suction issues when the doctor tried using her scope with the fms vue.External suction was used to allow proper suction of the fluid from the cavity.The case was completed using the same unit with no consequence to the patient.One fms vue will be returned for analysis by the customer.The tubing set was discarded.Additional information received by mitek complaint on (b)(6) 2014: the sales rep reported a delay over 30 minutes in the procedure.
|
The unit was evaluated and the reason for return : " inflow and outflow issues" could not be duplicated.The unit passed all diagnostic tests, functional tests, and is fully operational.Further, a review into the depuy mitek complaints system did not reveal any other previous complaints reported on this device.Based on the information provided, a root cause for the failure cannot be discerned.At this point, no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|