A medtronic representative reported that, while in a spinal fusion procedure, a site's suretrak2 clamp driver was visibly damaged, bent.Damage was discovered after the procedure, when attempting to remove the clamps from the instruments.The surgeon tightened down the clamps onto the instruments, causing the screws to strip and the driver to bend.No further details were provided.Replacement instrument was requested.There was no delay in therapy.The surgeon completed the procedure with the use of the navigation system.There was no impact on patient outcome.
|