• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER Back to Search Results
Model Number R2067
Device Problems Electrical /Electronic Property Problem (1198); Delamination (2904)
Patient Problem No Patient Involvement (2645)
Event Date 08/29/2014
Event Type  malfunction  
Event Description
It was reported that the radiofrequency programmer head originally returned as an associated device subsequently tested out of specification during manufacturer¿s analysis.There was no patient involvement.
 
Manufacturer Narrative
This report is based solely on device analysis.No information to suggest a device related adverse event or product problem was received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: during analysis the head failed uplink tests and was not able to interrogate and the cable was therefore replaced.It was further noted that the label's back coating was missing and the label was therefore replaced as well.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4312740
MDR Text Key5246001
Report Number2182208-2014-03533
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberR2067
Device Catalogue NumberR2067
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received10/08/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/22/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
2090 PROGRAMMER
-
-