MPRI SPRINT QUATTRO SECURE; DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 6947M62
Device Problems Device Contamination with Body Fluid (2317); Device Dislodged or Dislocated (2923)
Patient Problem No Code Available (3191)
Event Date 10/14/2014
Event Type  Injury  
Event Description
It was reported that within a week of implant, the right ventricular (rv) lead was found to have dislodged.The lead was attempted to be repositioned, but during the procedure, blood was found in the lead lumen, which in turn was leaking into the device header.The lead and device were explanted and replaced.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: the full lead was returned and analyzed.Both the outer insulation and the inner tubing of the lead were extrinsically breached due to a cut, and blood was found on the distal conductor of the lead and it was obstructed.The distal lv (low voltage) electrode of the lead was covered in blood and body tissue/fibrotic growth, and visual summary analysis of the lead indicated apparent explant damage.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE
Type of Device
DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4313333
MDR Text Key5207549
Report Number2649622-2014-14018
Device Sequence Number1
Product Code NVY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P920015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/01/2016
Device Model Number6947M62
Device Catalogue Number6947M62
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/14/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DVBB1D4 ICD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age00054 YR
Patient Weight116
-
-