• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problem Failure of Device to Self-Test (2937)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/22/2014
Event Type  malfunction  
Event Description
It was reported that the external pulse generator (epg) experienced a self-test error.The epg was returned to the manufacturer for servicing.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
 
Manufacturer Narrative
Product event summary: analysis confirmed the error code on the display, the main board was defective.As a result the main board was replaced.Further analysis was then performed on the main board after servicing of the device.Visual inspection revealed no anomalies.Bench analysis found that all tests passed on the automated test console.Additional analysis found that all circuitry operated normally.The log was then analyzed.During a scheduled battery/super cap measurement the super cap measured high which caused an error, the next power-up cycle then caused the displayed error to occur.Conclusion: the error was confirmed, however the cause of the error could not be determined.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4314934
MDR Text Key15103947
Report Number3004593495-2014-00069
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,company representati
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 10/22/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received11/06/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1356-2015
Patient Sequence Number1
-
-