Model Number 5392
Device Problem Failure of Device to Self-Test (2937)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/04/2014
Event Type  malfunction  
Event Description
It was reported that the external pulse generator (epg) displayed an error code.A functional check was requested.The epg was returned to the manufacturer for servicing.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.Product event summary: analysis confirmed the reported event.As a result the main printed circuit board (pcb) was replaced and calibrated.(b)(4).
 
Manufacturer Narrative
Further analysis was performed on the main board.Visual inspection revealed no anomalies.Bench analysis revealed no anomalies.The error log indicates that the displayed error was caused by a super capacitor high reading.Conclusion: the reported event was confirmed but no cause was determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 1190 0
MY  11900
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4316966
MDR Text Key5243317
Report Number3004593495-2014-00072
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/04/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received02/16/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-