The patient¿s generator was replaced due to end of service and was returned for product analysis.The reported ¿end of service¿ allegation was duplicated in the product analysis laboratory.The supply current tests did not meet functional specifications.These measurements demonstrate an increased current consumption for the device, potentially contributing to a premature end of life condition.A battery life estimation resulted in 4.60 years remaining before the eri flag would be set.However, an incomplete programming history with multiple gaps (4-year gap, 3-year gap) indicates the estimation does not use all the data required to make an accurate estimation.The increased current consumption was isolated to a leaky capacitor (c6).With the capacitor substitution for c6, the pulse generator module performed according to functional specifications.The most probable root cause for the premature end of life condition was identified to be a leaky capacitor, c6.The cause for the c6 capacitors increase in leakage could not be determined.
|