• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MINDRAY DS USA, INC. GAS MODULE III

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MINDRAY DS USA, INC. GAS MODULE III Back to Search Results
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/13/2014
Event Type  malfunction  
Event Description
Customer reported an issue with gas module iii, which may have affected gas monitoring.No pt injury was reported.
 
Manufacturer Narrative
Company rep evaluated the unit and verified the reported problem.Corrections included replacement of the unit's water trap housing, multi-gas module, and calibrating.Unit was calibrated and safety tested to factory's specs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GAS MODULE III
Type of Device
GAS MODULE
Manufacturer (Section D)
MINDRAY DS USA, INC.
mahwah NJ
Manufacturer Contact
wisara sethachutkul
800 macarthur blvd.
mahwah, NJ 07430
2019958045
MDR Report Key4319841
MDR Text Key17571930
Report Number2221819-2014-00941
Device Sequence Number1
Product Code BZK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062754
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Unknown
Type of Report Initial
Report Date 10/13/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/25/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/13/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-