Event Description
Customer reported an issue with the gas module iii, which may have affected gas monitoring.No pt injury was reported.
 
Manufacturer Narrative
Company representative evaluated the unit and verified the reported problem.Corrections included replacement of the unit's o2 filter tubing, readjusting the power supply voltage, and calibrating.Unit was calibrated and safety tested to factory's specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GAS MODULE III
Type of Device
GAS MODULE
Manufacturer (Section D)
MINDRAY DS USA, INC.
mahwah NJ
Manufacturer Contact
wisara sethachutkul
800 macarthur blvd.
mahwah, NJ 07430
2019958045
MDR Report Key4319855
MDR Text Key17581306
Report Number2221819-2014-00935
Device Sequence Number1
Product Code BZK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062754
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Unknown
Type of Report Initial
Report Date 10/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/25/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/15/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-