• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER COULTER LH 500 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER COULTER LH 500 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 178833
Device Problem Failure to Align (2522)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/14/2014
Event Type  malfunction  
Event Description
The customer reported that the manual sampling probe on the coulter lh 500 hematology analyzer appeared to be misaligned.There was no report of injuries, erroneous results, or change to patient treatment in connection with this event.A beckman coulter (bec) field service engineer (fse) was dispatched to the customer site to evaluate the instrument.
 
Manufacturer Narrative
The fse replaced the bent blood sampling valve to resolve the issue.The service was verified to meet the specified requirements per established procedures and the results meet published performance specifications.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 500 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s kraemer blvd
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
dung nguyen
250 s. kraemer blvd.
brea, CA 92821
7149614941
MDR Report Key4323084
MDR Text Key21939767
Report Number1061932-2014-02951
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042724
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number178833
Other Device ID NumberSOFTWARE VERSION 2A6
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/14/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-