MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. SJM MASTERS SERIES VALVE EXPANDED CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. SJM MASTERS SERIES VALVE EXPANDED CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 23AECJ-502
Device Problem Insufficient Information (3190)
Patient Problem Endocarditis (1834)
Event Date 11/14/2014
Event Type  Injury  
Event Description
The mechanical valve was explanted due to endocarditis.Another 23 mm sjm masters series valve was implanted.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
(b)(4).The results of this investigation concluded the leaflets opened and closed completely, and were freely mobile.There was a clinical history of prosthetic valve endocarditis, with no vegetations or acute inflammation seen.Special stains were negative for organisms.There was no evidence found to suggest the cause of the reported event was due to an intrinsic defect in the valve, as supported by review of the valve's device history record and the analysis performed.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES VALVE EXPANDED CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 0072 5
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 0072 5
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key4326147
MDR Text Key5241836
Report Number2648612-2014-00053
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
PP810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/17/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/20/2018
Device Model Number23AECJ-502
Device Catalogue Number23AECJ-502
Device Lot Number4000494
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/21/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age62 YR
-
-