• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ GMBH & CO. KG BIPOLAR CUTTING LOOP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ GMBH & CO. KG BIPOLAR CUTTING LOOP Back to Search Results
Model Number 26040GP1/6
Device Problem Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/01/2014
Event Type  malfunction  
Event Description
Allegedly, during a procedure a piece of the loop on the distal end broke off and fell into the patient; she immediately removed it.Procedure was completed and there was no serious injury to the patient.
 
Manufacturer Narrative
Evaluation of the instrument confirmed that a piece of the loop has broken off the distal end; we believe the break is possibly due to mechanical damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOLAR CUTTING LOOP
Type of Device
BIPOLAR CUTTING LOOP
Manufacturer (Section D)
KARL STORZ GMBH & CO. KG
tuttlingen
GM 
Manufacturer (Section G)
KARL STORZ GMBH & CO. KG
mittelstrasse 8
tuttlingen 7850 3
GM   78503
Manufacturer Contact
susie chen
2151 e. grand ave
el segundo, CA 90245-5017
4242188201
MDR Report Key4328638
MDR Text Key5243012
Report Number9610617-2014-00064
Device Sequence Number1
Product Code HIN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122983
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial
Report Date 11/04/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/05/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number26040GP1/6
Device Catalogue Number26040GP1/6
Device Lot Number47332
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-