BAXTER HEALTHCARE - IRVINE INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C1009KP
Device Problem Reflux within Device (1522)
Patient Problem No Patient Involvement (2645)
Event Date 11/10/2014
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This lot was manufactured from october 31, 2013 to november 1, 2013.Evaluation summary: the sample was received for evaluation.A visual inspection was performed with no issues noted.Functional flow testing was performed and the device was found to flow normally without leaks or backflow occurring.No malfunctions or abnormalities were identified during evaluation.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the reservoir of a large volume infusor had "reverse flow." this was noticed during or after filling and before patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4330964
MDR Text Key12818074
Report Number1416980-2014-45752
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
PMA/PMN Number
K071222
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Type of Report Initial,Followup
Report Date 11/20/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date10/01/2016
Device Catalogue Number2C1009KP
Device Lot Number13K009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2014
Date Manufacturer Received12/23/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-