• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PORTEX SPINAL ANESTHESIA TRAYS; CAZ - ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PORTEX SPINAL ANESTHESIA TRAYS; CAZ - ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number A3391-25
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Awareness during Anaesthesia (1707)
Event Type  Injury  
Event Description
A report was received that stated a patient received insufficient local anesthesia when the listed medical device was used.It was necessary to place patient under general anesthesia during the procedure.No adverse effects to patient reported.
 
Manufacturer Narrative
Field samples were not returned for evaluation.The supplier of the anesthesia medication performed a complete review of their manufacturing records for the lots that were reported to potentially be associated with the product issue.The supplier concluded that the product met with specifications at final testing.They also tested retained samples for potency of the same lots.All testing found the product to meet within specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX SPINAL ANESTHESIA TRAYS
Type of Device
CAZ - ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
keene NH
Manufacturer (Section G)
SMITHS MEDICAL ASD INC.
10 bowman dr.
keene NH 03431
Manufacturer Contact
pete hirte
1265 grey fox rd.
st. paul, MN 55112
6516287384
MDR Report Key4345000
MDR Text Key5141600
Report Number2183502-2014-00946
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K965017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/10/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberA3391-25
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA12/10/2014
Event Location Hospital
Date Manufacturer Received11/12/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-