• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - IRVINE INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - IRVINE INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C1009KP
Device Problem Reflux within Device (1522)
Patient Problem No Patient Involvement (2645)
Event Date 11/15/2014
Event Type  malfunction  
Event Description
It was reported that a large volume infusor" reversed flow".This malfunction occurred before use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device has been received by baxter, but the evaluation has not yet been completed.A follow-up report will be submitted upon completion of the evaluation or if any additional relevant information is received.
 
Manufacturer Narrative
(b)(4).The device was manufactured august 26, 2014 to august 27, 2014.Evaluation summary: the sample was received for evaluation.Visual inspection revealed that there was a leak/backflow from the fill port.Further visual inspection revealed a particle trapped under the checkband.Fourier transform infrared spectroscopy revealed that the particle was made of polyethylene.The cause of the backflow was determined to be the particle.A capa has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4347008
MDR Text Key5141637
Report Number1416980-2014-46639
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Type of Report Initial,Followup
Report Date 11/20/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date08/01/2017
Device Catalogue Number2C1009KP
Device Lot Number14H046
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2014
Date Manufacturer Received12/29/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-