Approximately six months post hvad implantation, the patient initially presented with low flows and electrical faults.A controller exchange was performed, but the e-faults persisted.After further inspection of the driveline, a splice repair was performed.Investigation is ongoing.
|
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Dob (b)(6) 1960.Adverse event.It was reported that the patient was taken by air ambulance to the implant center for repair of the driveline which he had inadvertently cut.Service report dated (b)(4) 2014: inspections showed damage in the outer sheath close to exit site.It was unclear if the patient or the site removed previous sheath repair.X-rays indicated potential recessed pins inside the driveline tube.It was reported on (b)(4) 2014, "faults are resolved.Patient doing well." x-rays- unknown dates.Preliminary log files - - normal power consumption operating on single stator.267 low flow alarms have been logged since november 14, 2014.The current low flow alarm limit is set to 1.0 lpm.2 vad disconnect alarms have been logged at the following times: (b)(6) 2014 at 19:12:34; (b)(6) 2014 at 12:26:07; 33 electrical fault alarms have been logged since (b)(6) 2014.Please note change in viscosity on (b)(6) 2014.Reviewed on (b)(4) 2014.Concomitant medical products: updated: controller catalog/part no.To 14063us.
|